Total Organic Carbon Analyzer

Installation

Teledyne Analytical Instruments

27

With the exception of the vent and carrier gas inlet, all other gas

connections are made on the same bottom panel as the liquid connections
and are shown in Figure 3-5. The NDIR vent and carrier gas connections
are made on the left side of the unit.

Gas Carrier Connections:

The facility carrier gas must be free of CO

2

and hydrocarbons and

be regulated to 15±2 psi and connected as shown in “As Built”
Drawings in the Appendix).

NDIR Vent (gas exhaust):

Exhaust connections must be consistent with the hazard level of the

constituent gases. Check Local, State, and Federal laws, and ensure that
the exhaust stream vents to an appropriately controlled area if required.

CAUTION: OPERATING THE UNIT WITHOUT CARRIER GAS

CAN DAMAGE THE ANALYZER.


3.5 Checking the System

Before applying power to the instrument, check:

• All gas and liquid lines as well as vent and drains are

properly connected and there are no leaks in the lines or at
fittings.

• Power and signal wiring are correct and that there are no

frayed or loose wires that could cause a short.

Prior to powering up the system and using the instrument for

analysis, the system must be configured for your application and
calibrated. This is described in Section 4.