Setup and Operation

Model 6750

Teledyne Analytical Instruments

36

The following window is illustrative of the D/DBPR [Run] screen.

4.3 Detailed Calibration Procedures

1. Although the instrument has been calibrated to your

specifications at the factory, it should be rechecked after
satisfactory installation and periodically calibrated as
suggested for your application.

2. Instrument calibration is performed by an “end-to-end”

method, whereby a known organic carbon chemical
standard solution is introduced to the analyzer and the
analyzer is “spanned” to that value.

3. Organic and Inorganic Carbon Standards:

a. (Potassium Hydrogen Phthalate) is recommended for

the organic carbon standard solution.

b. Sodium Carbonate is recommended for the inorganic

carbon standard.

c. Table 4-1 provides the concentration to be used for

two ranges of different chemical compounds (organic
& inorganic carbon).

d. Ratioing of these concentrations will provide other

ranges for the span solution.