Total Organic Carbon Analyzer

Theory of Operation

Teledyne Analytical Instruments

17

2.4.4 NDIR CO

2

Gas Detection

The key component for reliable TOC analysis is the precise and

interference-free detection of CO

2

. The Model 6750 NDIR unit requires

very little maintenance because of its ability to handle the corrosive
gases generated in TOC analysis. This is due in part to its non-reflective
borosilicate sample cell. It has no moving parts which further reduces
the maintenance load and it uses an internal self-calibrating technique.

The NDIR unit is shown in Figure 2-10. Some of the features of the

NDIR unit are as follows:

• Specific, interference-free CO

2

detection

• Dual-wavelength rationing compensates for drift

• Computer-controlled for accuracy

• Sapphire protected optics

• Non-corrosive, non-reflective sample cell (borosilicate)

• No moving parts or tools required for easy maintenance and

service

• No critical realignment required

Figure 2-10: Model 6750 NDIR Unit


The NDIR CO

2

detector uses a completely solid state, dual

wavelength ratioing technique with a single borosilicate sample cell that