Total Organic Carbon Analyzer

Maintenance

Teledyne Analytical Instruments

47

Figure 5-3: Installing Calibration Adapter

5. To set IR “Zero”, flow 200cc/minute oxygen (or CO

2

-

free air) for 5 minutes. Allow NDIR reading to stabilize.
Select [ZERO]. NDIR reading is reset to 0.

6. To set IR “Span” for the 6 inch (shorter) IR bench, use a

gas calibration standard mixture of 1% CO

2

in pure

nitrogen. For the longer (15 inch) bench, a gas mixture
of 0.1% CO

2

in pure nitrogen is required for. Flow 200

cc/minute of this cal gas for at least three minutes. Allow
NDIR reading to stabilize and then select [SPAN]. NDIR
reading is set to 10,000. Select [OK] to save or
[CANCEL] to reject the calibration setting.

The NDIR unit has now been properly calibrated. Reconnect the

fitting to reactor output tube.

5.3.2 NDIR Service

To remove and service the NDIR, simply follow the illustrated

sequence as described in this section following Figure 5-4.