Total Organic Carbon Analyzer

Theory of Operation

Teledyne Analytical Instruments

13

Figure 2-7: Sample Handling TOC-True

2.4.2.2 G

AS

P

HASE

S

AMPLE

H

ANDLING

Referencing Figure 3-5 and “As Built” drawings in the Appendix, a

carrier gas of (99.8%) pure oxygen or CO

2

and hydrocarbon-free air is

required. Erroneous TOC analysis will occur if the carrier gas has these
impurities, since both CO

2

and Hydrocarbon gases will contribute a

higher TOC value to the analysis than is actually present in the sample.

An Oxygen Generator (P/N ST200025) is available from Teledyne

that separates oxygen from the ambient air. It is a pressure swing
absorption PSA device requiring only electricity only to operate. Also
available from Teledyne is an Instrument Air Purifier (P/N ST200019).
This device can supply the oxygen demands of the instrument using
facility instrument air.

Facility carrier gas is introduced to the analyzer, regulated at 15± 2

psi, and then controlled by an internal mass flow controller (MFC-1) to
precisely control the flow rate of the carrier.

Note: Precise carrier gas flow control is mandatory for accurate