Maintenance

Model 6750

Teledyne Analytical Instruments

44

From this screen the operator may perform individual
testing of components related to functionality and
settings.

5.2 Troubleshooting

If you are experiencing trouble with your instrument, refer to Table

5-1: Troubleshooting.

Table 5-1: Troubleshooting

PROBLEMS

SOLUTION

Analyzer will not turn on

• Check fuse & power connections.
• Look for loose or damaged wiring.
• Check for 12V DC from power supply.
• Check power supply fuse.

Analyzer does not Calibrate, is
sluggish

• Look for & correct any liquid or gas line

leaks.

Analyzer is excessively noisy

• Check for leaks at tubing connections

Display not functioning

• Check cable connections
• Check power supply voltage
• Suspect:
• Malfunctioning Computer (P/N ST13039).
• Master Control Panel Assembly (P/N

ST13042-1)

Pump head not rotating

• Refer to section “Pumps” 12.3.12
• Verify that carrier loss did not cause the

pumps to be turned off.

• Check for AC voltage at terminal strip on

inside of application panel.

Pump head rotating but not
pumping fluid

• Refer to section “PUMPS”
• Change tubing
• Verify sample line is in liquid.