Total Organic Carbon Analyzer

Setup and Operation

Teledyne Analytical Instruments

39

Note: On the initial Startup, IR gas calibration should be

performed prior to liquid calibration

a. Select [CALIBRATE] from the main menu. The

following screen will appear:


b. Select [TOC].

The following screen appears:

c. Flow “ZERO” (D. I. Water) and allow NDIR reading

to stabilize. For maximum accuracy, especially at
lower ranges, this process may take 20 minutes.
When reading is stable, select [ZERO].

d. Flow Liquid Span Standard Solution. Allow NDIR

reading to stabilize, (approximately 15 minutes).

e. Select [SPAN]. It is recommended that the operator

continue to observe the reading and if not stabilized at
the desired span setting ± 2%, repeat the span
selection until stable.