beautypg.com

Teledyne 6750 - Total Organic Carbon analyzers User Manual

Page 49

background image

Total Organic Carbon Analyzer

Setup and Operation

Teledyne Analytical Instruments

37

Table 4-1: Preparation of Standards (TOC/TIC)

ORGANIC COMPOUNDS

100 mg/Liter

TOC (AS CARBON)

1000 mg/Liter

TOC (AS CARBON)

Ethylene Glycol

.233 ml/L H

2

0

2.33 ml/L H

2

0

Methanol

.337 ml/L H

2

0

3.37 ml/L H

2

0

Ethanol

.242 ml/L H

2

0

2.42 ml/L H

2

0

Acetone

.204 ml/L H

2

0

2.04 ml/L H

2

0

Carbon Tetrachloride

.807 ml/L H

2

0

8.07 ml/L H

2

0

Sucrose

.238 gm/L H

2

0

2.38 gm/L H

2

0

Urea

.500 gm/L H

2

0

5.00 gm/L H

2

0

Acetic Acid

.250 gm/L H

2

0

2.50 gm/L H

2

0

KHP (Potassium acid
phthalate)

.212 gm/L H

2

0

2.12 gm/L H

2

0

Glycine

.313 gm/L H

2

0

3.13 gm/L H

2

0

Sodium Stearate

.141 gm/L H

2

0

1.41 gm/L H

2

0

Succinic Acid

.246 gm/L H

2

0

2.46 gm/L H

2

0

Sodium Oxalate

.555 gm/L H

2

0

5.55 gm/L H

2

0

INORGANIC COMPOUNDS

50 mg/Liter

TIC (AS CARBON)

Sodium Carbonate

.442 gm/L H

2

0

Potassium Carbonate

.575 gm/L H

2

0

Ammonium Carbonate

.475 gm/L H

2

0

Add amounts shown (in milliliters or grams) to clean dry, one liter
volumetric flask and dilute to one (1) liter with distilled, deionized
water.

4. Reagent Preparation:


PREPARATION OF ONE MOLAR SODIUM
PERSULFATE SOLUTION (OXIDIZING REAGENT)