A-2

Appendix

Model 7120

Teledyne Analytical Instruments

7120 Specifications

% Range(s)

PPM Ranges

·

98-100% (CO2 purity)

0-500 ppm H2Ov in CL2, N2, CO or
COC/2

Other percent applications

Other ppm applications

·

95-100% CO in H2, N2, O2 0-1000ppm CO in flue gas

·

90-100% CH4 Purity

0-500 ppm N2O

·

0-1% H2Ov in Air

0-100ppm CO2 in Air

·

0-10% NH3 in Air

·

0-100% CH4 in Natural gas

·

0-10% CH4 in C2...C6 aliphatics

Above are typical percent and ppm ranges possible for a few sample gas

applications.

Other ranges and gases possible with optical cell path/optical filter changes
(consult factory).

Detector Type Lead salt as PbSe or PbS
Span Calibration
·

Calibration can be done in many ranges: choose a typical 80% of

range span gas.

Example: zero 99.995%CO2; span 98.40% CO2 in N2

·

Ultra pure Span calibration gas is recommended for purity applica-

tions.

Zero Calibration (Auto zero every 3 hours required)
·

Ultra pure Zero calibration to be performed using 99.99%

+

assay pure

CO2

·

Flowing reference gas required at 20 ccm (purity must be same as

ultra pure zero gas)

·

Non-purity applications can use sealed air reference

·

Zero gas from 0% level; use analyte free Air or N2

·

Span gas: use an 80 % full scale analyte in typically Air or N2 for the

pertinent range of interest.