Photometric Analyzer

Installation 2

2-13

Teledyne Analytical Instruments

Figure 2-8: FET Series Resistance

2.3.4

RS-232 Port

The digital signal output is a standard RS-232 serial communications

port used to connect the analyzer to a computer, terminal, or other digital
device. It requires a standard 9-pin D connector.

Output: The data output is status information, in digital form, updated

every two seconds. Status is reported in the following order:

The concentration in ppm or percent

The range in use (00 = Range 1, 01 = Range 2, 10 = Range 3, 11
= Range 4)

The span of the range (0-100 %, etc)

Which alarms—if any—are disabled (AL–x DISABLED)

Which alarms—if any—are tripped (AL–x ON).

Each status output is followed by a carriage return and line feed.

Input: The input functions using RS-232 that have been implemented

to date are described in Table 2-3.

Table 2-8: Commands via RS-232 Input

Command

Description

as

Immediately starts an autospan.

az

Immediately starts an autozero.

rp

Allows reprogramming of two System functions:
APPLICATION (gas use) and ALGORITHM (linearization).