3.8 AUTOMATIC FUNCTION

The events talking place during a zero cycle are as follows:

l. The zero cycle starts with activation of the calibration contact. The

analyzer outputs are held during the zero cycle.

2. The Auto Zero solenoid valve is activated and zero gas replaces the

sample. (This valve may be internal or external to the Model 7100

enclosure).

3. At about minute one in the cycle. The Auto Zero board then begins its

zeroing function, adjusting the processing circuitry of the analyzer to develop

a zero signal output. During this period, the analyzer display may swing

from below zero to above fullscale several times before it finally settles to

zero output. The “hold” feature prevents these wide output excursions from

affecting the analog output and trigger alarms. The Auto Zero circuit board

function takes about 6 minutes to complete. It is important that zero gas is

continually purging the cell during this time.

4. At around the 5th minute, the Auto Zero solenoid valve is deactivated and

sample replaces the zero gas.

5. The hold feature is deactivated and the display starts reading the sample

again, and the analog output soon starts tracking.

This concludes the zero cycle.

The time between zero cycles, may be programmed to last from one (1) to 23

hours.

For “Purity Measurements” typical zeroing is performed every 3 hours. This

may increase or decrease depending upon the complexity of the application

involved.