beautypg.com

Teledyne 7120 - NDIR analyzer User Manual

Page 31

background image

3 Start-up and Theory of Operation

Model 7120

3-8

Teledyne Analytical Instruments

NDIR Bench Output

Figure I

Gas Concentration

Figure II

Linearizer Input

Linearizer Output

to create a linear response. Refer to section 4.3.7 and 4.8.2 to see how

linearizer is programmed.

Piece-wise approximation is the method used to linearize the signal, i.e., the

linearizer’s output to input relationship can be graphed as a number of

straight line segments connected together to approximate the desired curve

that would be required to compensate for the nonlinearity of the bench

The points at which the compensation curve changes slope are called

breakpoints. The slope of each segment corresponds to the gain of the

linearizer in that segment, and this gain has to vary according to the input

voltage. This is achieved through software by adjusting a multiplier factor at

each straight line segment. The linearizer has nine specific voltage segments.

3.7 Control Unit

The Control Unit that controls the zero cycle and the sample system,

which may be common for all other Control Units is called the Slave Control

Unit. The master is unique because it contains the timer PC board and mode

switching programming for operation of any internal/external calibration or

auxillary valving.

The master control unit drives the other control units called slaves. The

slave control unit has all the features of the master except timing and control

of the sample system.

When a system contains the automatic calibration option, all control

units are slaves. The Auto-Cal function controls the sample system and the

zero cycle and provides a span cycle.