Photometric Analyzer

Installation 2

2-9

Teledyne Analytical Instruments

Figure 2-4: Alarm Relay Contact Pins

THRESHOLD

ALARM 1

Press here to

insert wire.

Release to hold.

Insert wire

here.

Moving contact

Normally open

Normally closed

Remote Calibration Protocol: To properly time the Digital Remote

Cal Inputs to the Model 7120 Analyzer, the customer's controller must
monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote Cal

Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It

will ignore your request to calibrate, and it will not remember that request.

Once a zero or span command is sent, and acknowledged (contact

closes), release it. If the command is continued until after the zero or span is
complete, the calibration will repeat and the Cal Relay Contact (CRC) will
close again.

For example:

1) Test the CRC. When the CRC is open, Send a zero command

until the CRC closes (The CRC will close quickly.)

2) When the CRC closes, remove the zero command.
3) When CRC opens again, send a span command until the CRC

closes. (The CRC will close quickly.)

4) When the CRC closes, remove the span command.