4 Operation

Model 7120

4-10

Teledyne Analytical Instruments

The self diagnostics are run automatically by the analyzer whenever the

instrument is turned on, but the test can also be run by the operator at will. To
initiate a self diagnostic test during operation, use the UP/DOWN key to scroll
through the MAIN MENU to the

SELF–TEST and Enter. The screen will follow

the running of the diagnostic.

RUNNING DIAGNOSTIC
Testing Preamp — Cell

When the testing is complete, the results are displayed.

Power: OK Analog: OK
Cell: 2 Preamp: 3

The module is functioning properly if it is followed by

OK. A number

indicates a problem in a specific area of the instrument. Refer to Chapter 5
Maintenance and Troubleshooting
for number-code information. The results
screen alternates for a time with:

Press Any Key
To Continue...

Then the analyzer returns to the initial System screen.

4.3.6 The Model Screen

Scroll through the MAIN MENU to MODEL and Enter. The screen

displays the manufacturer, model, and software version information.

4.3.7 Checking Linearity with ALGORITHM

Use UP/DOWN control to select

ALGORITHM, and Enter.

sel rng to set algo:
—> Ш1 Ш2 Ш3 <—

Use the UP/DOWN Control to select the range: 01, 02, or 03. Then press

Enter

.

Gas Use: CO2
Range: Ø — 10%

Enter again.

Algorithm setup:
VERIFY SET UP

Use UP/DOWN key to select and Enter

VERIFY to check whether the

linearization has been accomplished satisfactorily.