beautypg.com

0 start-up and operation – Teledyne 7120 - NDIR analyzer User Manual

Page 24

background image

Photometric Analyzer

Start-up and Operation

3-1

Teledyne Analytical Instruments

3.0 Start-up and Operation

3.1 Preliminary

Before applying power to the system, TAI suggests that the electrical

wiring installation be checked against the system input-output diagram.

Proper attention to this preliminary check will prevent severe damage caused

by wiring errors.

Also, verify that all connections to the system have been made

correctly. Refer to the system outline diagram for proper connections.

3.2 NDIR Analyzer Startup

Power up the unit by depressing the rear panel switch. From a first

time power-on cold start attempt, allow (1) one hour warm-up to proceed.

Observe that the Digital display will go through a diagnostic routine before

the readings revert to a continuous concentration readout.

3.2.1 Initial Set-up and Zeroing

Assure the sample will enter from the zero inlet gas position. Open the

zero gas tank and set the pressure regulator to 20 psig. Set the zero gas flow

to 2 SCFH (1 lpm). Set the sample flow to 2 SCFH (1 lpm). When

applicable, assume the flowing reference cell gas is flowing between 20-50

ccm.

Initialize a zero operation through the system menu. Refer to Section 4

for Electronics /Control Unit Modes and Functioning to navigate the zero

menu.

After the zero cycle is over, the instrument is automatically brought in

the sampling cycle. An inaccurate reading of the concentration of the gas of

interest will be obtained. The instrument is not yet calibrated. Permit the

instrument to operate in this mode for at least one hour to stabilize if not

already done so upon initial power up.

NOTE: The following options could be included in your system:

In case the instrument is part of a multi instrument system, where all

instruments are connected to the same sample system and under control of a

single timer, all instruments of the system will go through the zero and

sample cycle simultaneously. The Control Unit of the instrument which