0 maintenance – Teledyne 7120 - NDIR analyzer User Manual

Page 68

background image

Infrared Analyzer

Maintenance 5

5-1

Teledyne Analytical Instruments

5.0 Maintenance

Aside from normal cleaning and checking for leaks at the gas connec-

tions, routine maintenance is limited to replacing filter elements and fuses,
and recalibration.

WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS

MANUAL.

5.0

Replacing the Fuse

Remove Power to Unit before replacing the
fuse.

When a fuse blows, check first to determine the cause, then replace the

fuse using the following procedure:

1. Disconnect the AC power and place the power switch located on

the rear panel in the O position. Remove the power cord from the
receptacle.

2. The fuse receptacle is located in the power cord receptacle

assembly in the upper left-hand corner of the rear panel. See
Figure 5-1

Figure 5-1: AC Fuse Replacement

FUSE