beautypg.com

Teledyne 7120 - NDIR analyzer User Manual

Page 51

background image

4 Operation

Model 7120

4-18

Teledyne Analytical Instruments

4.4.1.2

Manual Mode Zeroing

Scroll to Zero and enter the Zero function. The screen that appears allows

you to select between automatic or manual zero calibration. Use the UP/DOWN
keys to toggle between

AUTO and MAN zero settling. Stop when MANUAL

appears, blinking, on the display.

Select zero
mode: MANUAL

Enter to begin the zero calibration. After a few seconds the first of three

zeroing screens appears. The number in the upper left hand corner is the first-
stage zero offset. The microprocessor samples the output at a predetermined
rate.

####.## % CO2
Zero adj:2048 C—Zero

The analyzer goes through C–Zero, F–Zero, and S–Zero. During C–Zero

and F–Zero, use the UP/DOWN keys to adjust displayed

Zero adj: value as

close as possible to zero. Then, press Enter.

S–Zero starts. During S–Zero, the Microcontroller takes control as in Auto

Mode Zeroing, above. It calculates the differences between successive sam-
plings and displays the rate of change as

Slope= a value in parts per million per

second (ppm/s).

####.## % CO2
Slope=#.### S—Zero

Generally, you have a good zero when

Slope is less than 0.05 ppm/s for

about 30 seconds.

Once zero settling completes, the information is stored in the analyzer’s

memory, and the instrument automatically returns to the Analyze mode.

4.4.1.3 Cell Failure

Cell failure in the 7100 is usually associated with inability to zero the

instrument with a reasonable voltage differential across the Wheatstone bridge. If
this should ever happen, the 7100 system alarm trips, and the VFD displays a
failure message.

Cell cannot be balanced
Check your zero gas

Before replacing the sensor:

a. Check your zero gas to make sure it is within specifications.