beautypg.com

Teledyne 7120 - NDIR analyzer User Manual

Page 25

background image

3 Start-up and Theory of Operation

Model 7120

3-2

Teledyne Analytical Instruments

houses the timer and operates the sampling system is called the master. The

other Control Units of the other instruments are called slaves.

Each of the other instruments may be monitoring the concentration of

several different gases of interest in the sample, for example CO, CO2 and/or

Combustibles as CH4 in flue gas.

3.2.2 Operational Calibration

After the instrument has stabilized, let zero gas flow through the

analyzer

Perform a zero of the analyzer.

Perform a span of the analyzer.

Open the span gas tank and set the pressure regulator to 20 psig.

Switch the mode switch to sample. Refer to Section 4 again.

Induce an automatic zero cycle as described in Section 4 also.

After the zero cycle, the analyzer reverts to the sample cycle. The

sample reading is now accurate and the analyzer is placed in continuous

operation. See Sections 4.6-4.8.

NOTE: In case slave analyzers are involved, calibrate them

simultaneously with the master analyzer.

After the instrument is calibrated, when no Auto-Cal option was

selected, shut off the main valve on the span gas tank. This tank is not used

during automatic sampling. Leave the zero tank open.

3.3

Theory of operation

3.3.1 General

T

he non-dispersive infrared (NDIR) analyzer is one of the major components

of the system. It employs the basic principles of spectroscopic analysis to

measure a specific concentration of one gas in a multicomponent gas system.

The concentration of a gas is determined by exposing a chamber filled with a

gas mixture to infrared radiant energy and measuring how much of the

specific (non-dispersive) infrared wavelength is absorbed by the gas being

measured.

As an example, the NDIR analyzer is used most in flue gas applications

where the amount of carbon monoxide in a flue gas mixture is measured.

The specific infrared wavelength at which the carbon monoxide molecule

absorbs infrared energy is at 4.65 microns. The more carbon monoxide

present in the measurement cell, the more energy its molecules absorb.

The NDIR analyzer needs four basic components to measure the spectral