Teledyne 3350 - Microprocessor based control room monitor for personnel safety User Manual
Page 43
A-7
Alarm Oxygen Monitor
Appendix
Teledyne Analytical Instruments
Section V – Emergency and First Aid Procedures
NOTE: The oxygen sensors are sealed, and under normal circumstances,
the contents of the sensors do not present a health hazard. The
following information is given as a guide in the event that a cell
leaks.
Protective clothing:
Rubber gloves, chemical splash goggles.
Clean-up procedures:
Wipe down the area several times with a wet pa-
per towel. Use a fresh towel each time.
Protective measures
during cell replacement:Before opening the bag containing the sensor
cell, check the sensor cell for leakage. If the sen-
sor cell leaks, do not open the bag. If there is
liquid around the cell while in the instrument,
put on gloves and eye protection before remov-
ing the cell.
Disposal:
Should be in accordance with all applicable
state, local and federal regulations.
NOTE: The above information is derived from the MSDS provided by the
manufacturer. The information is believed to be correct but does
not purport to be all inclusive and shall be used only as a guide.
Teledyne Analytical Instruments shall not be held liable for any
damage resulting from handling or from contact with the above
product.
Section VI – Handling Information
Eye Contact:
Flush eyes with water for at least 15 minutes and get im-
mediate medical attention.
Skin Contact:
Wash affected area with plenty of water and remove
contaminated clothing. If burning persists, seek medical
attention.
Ingestion:
Give plenty of cold water. Do not induce vomiting.
Seek medical attention. Do not administer liquids to an
unconscious person.
Inhalation:
Liquid inhalation is unlikely.