3 Installation

Model 3350

3-6

Teledyne Analytical Instruments

3.5.3

Analog Outputs

There are four DC output signal connectors with screw terminals on the

panel. Recommended wire: 22 AWG minimum with 300VAC insulation.
There are two wires per output with the polarity noted. The outputs are:

0–10 V % Range:

Voltage rises with increasing oxygen concentration,
from 0 V at 0 percent oxygen to 10 V at full scale
percent oxygen. (Full scale = 25% O

2

).

0–10 V Range ID:

03.33 V = Low Range, 06.66 V = High Range,
10 V = Air Cal Range.

4–20 mA % Range:

Current increases with increasing oxygen concentra-
tion, from 4 mA at 0 percent oxygen to 20 mA at full
scale percent oxygen. (Full scale = 25% O

2

).

Audible Alarm:

Factory installed to internal buzzer, may be used for
customer interconnection to 12V a Buzzer (12-
15VDC, 4.3mA Max.).

The RS232 terminals must not be connected to
any AC line source, or any source of hazardous
voltage.

3.5.4 RS-232 Port (Optional)

The digital signal output is a standard RS-232 serial communications

port used to connect the analyzer to a computer, terminal, or other digital
device. The pinouts are listed in Table 3-1.

Table 3-1: RS-232 Signals

RS-232 Sig

RS-232 Pin

Purpose

RD

2

Received Data

TD

3

Transmitted Data

COM

5

Common

The data sent is status information, in digital form, updated every two

seconds. Status is reported in the following order:

The concentration in percent

The range in use