beautypg.com

Teledyne 3350 - Microprocessor based control room monitor for personnel safety User Manual

Page 27

background image

3-11

Alarm Oxygen Monitor

Installation 3

Teledyne Analytical Instruments

7. Any and all necessary warning labels, notices, and information

are provided.

8. Operators understand the operations and functions of the

analyzer and system.

9. Any environmental or other influences that could affect the

operation or accuracy of the instrument are identified and
handled.

3.10 Accessory Flow-Through Adapter

A flow-through adapter is available for the Series 3350

Oxygen Analyzer for those applications that require piped-in gases. The
adapter consists of a sealed chamber where the instrument's probe is
inserted, with two radially-oriented hose connectors to which supply and
vent lines for the calibration gas are connected. Refer to Fig 3-7 . The
design provides gas flow over the sensing surface of the cell without
contamination by surrounding monitored atmosphere.

NOTE: If the sensor probe is installed without a constant flow

of gas, it will interfere with the sensor readings and will cause drift
and inaccurate readings.

Fig. 3-7 Flow-through Adapter and sensor Probe