A-1

Percent Paramagnetic Oxygen Analyzer

Appendix

Teledyne Analytical Instruments

Appendix

A-1 Specifications

Packaging: Explosion-proof. Bulkhead mount.

Sensor: Paramagnetic Sensor.

Ranges: Three user definable ranges from 0-5 % to 0-

100 %.

Autoranging with range ID output.

Sample System: Flow indicator visible from front of unit.

Positive pressure service.

Vacuum service (optional).

Auto Cal / Auto Zero. (Available with op-
tional, electrically operated valves.)

Alarms: One system-failure alarm contact to detect

power failure.

Two adjustable concentration threshold alarms
with fully programmable setpoints.

Diagnostics: Start-up or on-demand self-testing function.

Displays: 2 line by 20 alphanumeric, VFD screen, and

one 5 digit LED display.

Digital Interface: Full duplex RS-232 communications port.

Power: Universal power supply 100-240 V ac, at 50

or 60 Hz, 70 Watts max.

Operating Temperature: 0-45 °C

Accuracy: ±2% of full scale at constant temperature.

±5% of full scale over operating temperature