3 sensor replacement, 4 fuse replacement – Teledyne 3020M - Explosion proof paramagnetic oxygen analyzer User Manual

Page 56

background image

5 Maintenance

Model 3020M

5-2

Teledyne Analytical Instruments

Figure 5-1: Major Internal Components

To swing open the cover panel, remove all screws.

5.3

Sensor Replacement

The Paramagnetic Sensor is maintenance free, and has a long life.

Should the sensor fail, contact the factory for service and replacement.

5.4

Fuse Replacement

The 3020M requires two 5 x 20 mm, 4 A, T type (Slow Blow) fuses.

The fuses are located inside the explosion proof housing on the Electrical
Connector Panel, as shown in Figure 5-2. To replace a fuse:

1. Disconnect the Unit from its power source.

2. Place a small screwdriver in the notch in the fuse holder cap,

push in, and rotate 1/4 turn. The cap will pop out a few
millimeters. Pull out the fuse holder cap and fuse, as shown in
Figure 5-3.

Preamp PCB

Main PCB

Display PCB

Sensor

Gas Control Valves

Customer Interface Panel
(Electrical Connector Panel)

Door Subassembly

(Cross Section)

Box Subassembly

(Gas Connector Panel
not shown)

24V Sensor Power Supply
located behind Customer
Interface Board.