Teledyne 3020M - Explosion proof paramagnetic oxygen analyzer User Manual

Page 52

background image

4 Operation

Model 3020M

4-16

Teledyne Analytical Instruments

The automatic air calibration range is always 0-25 % and is not pro-

grammable.

4.11.1 Setting the Analog Output Ranges

To set the ranges, ENTER the

RANGE

function mode by selecting

RANGE

from the MAIN MENU. The

RANGE

screen appears.

L—5.00 M—25.00
H—100.00 Mode—AUTO

Use the DOWN/UP control to the range to be set: low (

L

), medium (

M

),

or high (

H

).

Use the DOWN/UP

control to enter the upper value of the range (all

ranges begin at 0 %). Repeat for each range you want to set. ENTER to
accept the values and return to the Analysis Mode. (See note below.)

Note: The ranges must be increasing from low to high, for example,

if range 1 is set as 0–10 % and range 2 is set as 0–100 %, range
3 cannot be set as 0–25 % since it is lower than range 2.

4.11.2 Automatic Ranging

After defining your analysis ranges, set

Mode–

to

AUTO

, if not already

there. ESCAPE to re-enter the ANALYZE screen using the fixed range.

4.11.3 Fixed Range Analysis

The autoranging mode of the instrument can be overridden, forcing the

analyzer DC outputs to stay in a single predetermined range.

To switch from autoranging to fixed range analysis, ENTER the

RANGE

function by selecting

RANGE

from the MAIN MENU.

Use the DOWN/UP

control to move

AUTO

between the arrows.

Use the DOWN/UP

control to switch from

AUTO

to

FX/LO

,

FX/

MED,

or

FX/HI

to set the instrument on the desired fixed range (low, me-

dium, or high).

L—5.00 M—25.00
H—100.00 Mode—FX/LO

or

L—5.00 M—25.00
H—100.00 Mode—FX/MED

or