3 Installation

Model 3020M

3-10

Teledyne Analytical Instruments

DTR

4

Data Terminal Ready

COM

5

Common

DSR

6

Data Set Ready

RTS

7

Request to Send

CTS

8

Clear to Send

RI

9

Ring Indicator

The data sent is status information, in digital form, updated every two

seconds. Status is reported in the following order:

The concentration in percent

The range in use (HI, MED, LO)

The span of the range (0-100 %, etc)

Which alarms—if any—are disabled (AL–x DISABLED)

Which alarms—if any—are tripped (AL–x ON).

Each status output is followed by a carriage return and line feed.

Four input functions using RS-232 have been implemented to date.

They are described in Table 3-4.

Table 3-4: Commands via RS-232 Input

Command

Description

as

Immediately starts an autospan.

az

Immediately starts an autozero.

st

Toggling input. Stops/Starts any status message output from
the RS-232, until st is sent again.

The RS-232 protocol allows some flexibility in its implementation.

Table 3-5 lists certain RS-232 values that are required by the 3020M imple-
mentation.

Table 3-5: Required RS-232 Options

Parameter

Setting

Baud

2400

Byte

8 bits

Parity

none

Stop Bits

1

Message Interval

2 seconds