Percent Paramagnetic Oxygen Analyzer
Operation 4
4-17
Teledyne Analytical Instruments
L—5.00 M—25.00
H—100.00 Mode—FX/HI
ESCAPE to re-enter the ANALYZE screen using the fixed range.
Note: When performing analysis on a fixed range, if the oxygen
concentration rises above the upper limit, as established by
the setup, for that particular range, the output saturates at 1
V dc (20 mA at optional mA dc output). However, the digital
readout and the RS-232 output continue to read the true value
of the oxygen concentration regardless of the analog output
range.
4.12 The CONTRAST Function
If you cannot read anything on the display after first powering up:
1. Observe LED readout.
a. If LED meter reads all eights and dots, go to step 3.
b. If LED meter displays anything else, go to step 2.
2. Disconnect power to the Analyzer and reconnect again. LED
meter should now read all eights and dots.
4.13 The STANDBY Function
In
STANDBY
, the analyzer’s internal circuits are powered, but there
are no displays or outputs from the analyzer.
WARNING: THE POWER CABLE MUST BE UNPLUGGED TO
FULLY DISCONNECT POWER FROM THE INSTRU-
MENT. WHEN THE ACCESS DOOR IS OPEN AND
THE POWER CABLE IS CONNECTED, EXTRA
CARE IS REQUIRED TO AVOID CONTACT WITH
LIVE ELECTRICAL CIRCUITS.
CAUTION:
If you disconnect the primary power source from
the analyzer, then on re-energizing, you will be
required to choose to keep the configuration you
previously programmed into your instrument in the
Setup Mode, or it will reset to factory defaults.
Contrast Function is
DISABLED
(Refer to Section 1.6)