beautypg.com

40 % of n, And 50% of ar. c0 – Teledyne 3020M - Explosion proof paramagnetic oxygen analyzer User Manual

Page 18

background image

2-4

2 Operational Theory

Model 3020M

Teledyne Analytical Instruments

Gas

Cross

Gas

Cross

Sensitivity

Sensitivity

in vol. %

in vol. %

Acetylene C

2

H

2

-0.24

Hydrogen chloride HCI

-0.30

Allene C

3

H

4

-0.44

Hydrogen fluoride HF

+0.10

Ammonia NH

3

-0.26

Hydrogen iodide HI

-1.10

Argon Ar

-0.22

Hydrogen sulphide H

2

S -0.39

Bromine Br

2

-1.30

Iodine I

-2.40

1.2 Butadiene C

4

H

6

-0.49

Isobutane C

4

H

10

-1.11

1.3 Butadiene C

4

H

6

-0.49

Isopantane C

5

H

12

-1.49

n-Butane C

4

H

10

-1.11

Krypton Kr

-0.51

I-Butane C4H8

-085

Laughing gas N

2

O

-0.20

cis 2-Butane C

4

H

8

-0.89

Methane CH

4

-0.20

trans 2-Butane C

4

H

8

-0.92

Neon Ne

+0.13

Carbon dioxide CO

2

-0.27

Neopentane C

5

H

12

-1.49

Carbon monoxide CO

+0.06

Nitric acid HNO

3

+0.43

Chlorine Cl

2

-0.77

Nitrogen dioxide NO

2

+28.00

Cyclo hexane C

6

H

12

-1.56

Nitrous oxide NO

+40.00

Ethane C

2

H

6

-0.43

n-Octane C

8

H

18

-2.50

Ethylene C

2

H

4

-0.26

n-Pentane C

5

H

12

-1.45

Helium He

+0.30

Propane C

3

H

8

-0.86

n-Heptane C

7

H

16

-2.10

Propylene C

3

H

6

-0.55

n-Hexane C

6

H

14

-1.70

Vinyl chloride

-0.63

Hydrogen H

2

+0.24

Water H

2

O

-0.02

Hydrogen bromide HBr

-.0.61

Xenon Xe

-0.95

With gas mixtures the components are weighted according to their

proporational volumes.

The cross interference with a gas mixture can be determined in advance if the

concentration of the individual background gases is known.

Example:

Cross sensitivity calculation for a gas mixture:

The gas to be measured consists of 10 % CO

2

, 40 % of N

2

, and 50%

of Ar.

C0

2

: -0.27 x 10% = -0.027

Ar: -0.22x5O%= -0.11

TOTAL CROSS INTERFERENCE -0.137

Calculation of the zero value: