3 Installation

Model 3020M

3-6

Teledyne Analytical Instruments

Table 3-1: Analog Concentration Output—Example

Voltage Signal

Current Signal

% O

2

Output (V dc)

Output (mA dc)

0

0.0

4.0

1

0.1

5.6

2

0.2

7.2

3

0.3

8.8

4

0.4

10.4

5

0.5

12.0

6

0.6

13.6

7

0.7

15.2

8

0.8

16.8

9

0.9

18.4

10

1.0

20.0

To provide an indication of the range, a second pair of analog output

terminals are used. They generate a steady preset voltage (or current when
using the current outputs) to represent a particular range. Table 3-2 gives the
range ID output for each analysis range.

Table 3-2: Analog Range ID Output—Example

Range

Voltage (V)

Current (mA)

LO

0.25

8

MED

0.50

12

HI

0.75

16

0-25%

Non Programmable

1.00

20

NOTE:

Impendance across 4-20mA output should not exceed 1000
ohms. Note polarity of terminals when wiring to the analog
outputs. Do not short 0-1VDC output for long period of time,
or damage may occur to analyzer.

CAUTION: Voltage may be present on alarm contacts when line power

is disconnected.