Teledyne 3020M - Explosion proof paramagnetic oxygen analyzer User Manual
Page 48
4 Operation
Model 3020M
4-12
Teledyne Analytical Instruments
If you have trouble zeroing, you may need to skip to section 4.8.1.3
Cell Failure.
Auto Mode Zeroing
Select
ZERO
to enter the
ZERO
function. The
ZERO
screen allows
you to select whether the zero calibration is to be performed automatically or
manually. Use the DOWN/UP control to toggle between
AUTO
and
MAN
zero settling. Stop when
AUTO
appears on the display.
Zero: Settling: AUTO
ENTER to begin zeroing.
#### % Zero
Slope=#### ppm/s
The beginning zero level is shown in the upper left corner of the dis-
play. As the zero reading settles, the screen displays and updates information
on
Slope
(unless the Slope starts within the acceptable zero range and does
not need to settle further).
Then, and whenever Slope is less than 0.08 for at least 3 minutes,
instead of Slope you will see a countdown, such as
1 Left
,
0 Left
. These are
steps in the zeroing process that the system must complete, AFTER settling,
before it can go back to ANALYZE.
#### % Zero
1 Left=### ppm/s
The zeroing process will automatically conclude when the output is
within the acceptable range for a good zero. Then the analyzer automatically
returns to the ANALYZE screen.
Manual Mode Zeroing
ENTER the
ZERO
function. The screen that appears allows you to
select between automatic or manual zero calibration. Use DOWN/UP to
toggle between
AUTO
and
MAN
zero settling. Stop when
MAN
appears on
the display.
Zero: Settling: Man
ENTER to begin the zero calibration. After a few seconds the first of
five zeroing screens appears. The number in the upper left hand corner is the
first-stage zero offset. The microprocessor samples the output at a predeter-
mined rate. It calculates the differences between successive samplings and