Percent Paramagnetic Oxygen Analyzer

Maintenance 5

5-3

Teledyne Analytical Instruments

Figure 5-2: Removing Fuse Cap and Fuse from Holder

2. Replace fuse by reversing process in step 1.

5.5

System Self Diagnostic Test

Use the

DOWN/UP

control

to scroll through the MAIN MENU to

SELF-TEST

. The screen will follow the running of the diagnostic.

RUNNING DIAGNOSTIC
Testing Preamp — 83

During preamp testing there is a countdown in the lower right corner of

the screen. When the testing is complete, the results are displayed.

Power: OK Analog: OK
Preamp: 3

The module is functioning properly if it is followed by

OK

. A number

indicates a problem in a specific area of the instrument. Refer to Table 5-1
for number-code information. The results screen alternates for a time with:

Press Any Key
To Continue...

The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power

0

OK

1

5 V Failure

2

15 V Failure

3

Both Failed