1 Introduction

Model 3020M

Teledyne Analytical Instruments

1.5.1 UP/DOWN Switch

Functions: The UP/DOWN switch is used to select the function to be

performed. Choose UP or DOWN to scroll through the following list of
twelve functions:

Auto-Cal Set up an automatic calibration sequence.

PWD

Install a password to protect your analyzer setup.

Logout

Locks Setup Mode.

Version

Displays model and version of analyzer.

Self-Test Runs internal diagnostic program, displays results.

Span

Span calibrate the analyzer.

Zero

Zero calibrate the analyzer.

Alarms

Set the alarm setpoints and attributes.

Range

Set up the 3 user definable ranges for the instrument.

Contrast Adjust LCD contrast.

Standby

Leave analyzer powered, but no outputs or displays.

WARNING: THE POWER CABLE MUST BE DISCONNECTED TO

FULLY REMOVE POWER FROM THE INSTRUMENT.

CAUTION: VOLTAGE MAY BE PRESENT ON ALARM CON-

TACTS WHEN LINE POWER IS DISCONNECTED.

Subfunctions: Once a Function is entered, the UP/DOWN switch is

used to select between any subfunctions displayed on the VFD screen.

Parameter values: When modifiable values are displayed on the

VFD, the UP/DOWN switch can be used to increment or decrement the
values.

1.5.2 ESCAPE/ENTER Switch

Data Entry: The ESCAPE/ENTER switch is used to input data, from

the alphanumeric VFD screen into the instrument:

Escape Moves VFD display back to the previous screen in a

series. If none remains, returns to the

Analyze

screen.

With subfunction selected, moves VFD back through
items on screen, to first item, then moves VFD to
previous display.

Enter

With a Subfunction Selected: Moves VFD on to the
next screen in a series. If none remains, returns to the

Analyze

screen.

Contrast Function is

DISABLED

(Refer to Section 1.6)