beautypg.com

Appendix – Teledyne 3000TA-EU - General purpose trace oxygen analyzer User Manual

Page 79

background image

Trace Oxygen Analyzer

Appendix

Teledyne

Analytical

Instruments

67

Appendix

A-1 Model 3000TA Specifications

Packaging: General Purpose

 Flush panel mount (Standard).
 Relay rack mount. Contains either

one or two instruments in one 19"
relay rack mountable plate
(Optional).

Sensor: Teledyne L-2 trace analysis

Micro-Fuel Cell. Cell Block: 316 stainless
steel.

Sample System: All wetted parts of 316 stailess steel. 90 %

Response Time: 65 seconds at 25 °C (77
°F).

Ranges: Three user definable ranges from 0–10

ppm to 0–250,000 ppm, plus air
calibration range 0- 250,000 ppm (25 %).

Autoranging with range ID output.

Alarms: One system-failure alarm contact to detect

power failure or sensor-zero failure.

Two adjustable concentration threshold

alarm contacts with fully programmable
setpoints.

Displays: Two-line by 20-character, VFD screen,

and one 5 digit LED display.

Digital

Interface: Full duplex RS-232 communications port.

Power: Universal power supply 85-250 VAC, at

47-63 Hz, 0.9 A MAX.

Operating Temperature: 0-50 °C (32-122 °F)
Humidity: 99% Altitude: 1,609 m

Accuracy: ±2% of full scale at constant temperature.

±5% of full scale over operating