Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

19

3.3 Rear Panel Connections

Figure 3-3 shows the Model 3000TA rear panel. There are ports for

gas, power, and equipment interface. The Zero In and Span In ports are
not included on the standard model, but are available as options.

Figure 3-3: Rear Panel of the Model 3000TA

3.3.1 Gas Connections

Before using this instrument, it should be determined if the unit will

be used for pressurized service or vacuum service and low pressure
applications. Inspect the restrictor kit that came with the unit. The kit
consist of two restrictors and a union for 1/4” diameter tubing. Notice
that the two 1-3/4” long, 1/4” diameter tubing are restrictors. It has an
open end and a closed end with a small circular orifice. The restrictor
without the blue sticker is for low pressure and vacuum service. For high
pressure applications (5-50psig), use the restrictor that has a blue sticker
on the body.

For pressurized service (> 5psig) , use the restrictor with the blue

dot and union from the restrictor kit and attach it to the Sample In port.