Maintenance 3000TA-

EU

Teledyne

Analytical

Instruments

62

2. To change between American and European fuses, remove

the single retaining screw, flip Fuse Block over 180 degrees,
and replace screw.

3. Replace fuse as shown in Figure 5-3.
4. Reassemble Housing as shown in Figure 5-2.

American Fuses

European Fuses

Figure 5-3: Installing Fuses

5.4 System Self Diagnostic Test

1. Press the System button to enter the system mode.
2. Use the ◄►arrow keys to move to More, and press Enter.
3. Use the ◄►arrow keys to move to Self-Test, and press

Enter

. The following failure codes apply:

Table 5-1: Self Test Failure Codes

Power
0 OK
1

5 V Failure

2

15 V Failure

3 Both

Failed

Analog
0 OK
1

DAC A (0–1 V Concentration)

2

DAC B (0–1 V Range ID)

3 Both

Failed