Installation 3000TA-

EU

Teledyne

Analytical

Instruments

26

Note: The Remote Valve connections (described below) provides

signals to ensure that the zero and span gas valves will be
controlled synchronously. If you have the –C Internal valve
option—which includes additional zero and span gas
inputs— the 3000T automatically regulates the zero, span
and sample gas flow.

Range ID Relays: Four dedicated Range ID relay contacts. The

first three ranges are assigned to relays in ascending order—Low range
is assigned to Range 1 ID, Medium range is assigned to Range 2 ID, and
High range is assigned to Range 3 ID. The fourth range is reserved for
the Air Cal Range (25%). Table 3-4 lists the pin connections.

Table 3-4: Range ID Relay Connections

Pin Function
21

Range 1 ID Contact

38

Range 1 ID Contact

22

Range 2 ID Contact

39

Range 2 ID Contact

19

Range 3 ID Contact

18

Range 3 ID Contact

34

Range 4 ID Contact (Air Cal)

35

Range 4 ID Contact (Air Cal)


Network I/O:
A serial digital input/output for local network

protocol. At this printing, this port is not yet functional. It is to be used
for future options to the instrument. Pins 13 (+) and 29 (–).


Remote Valve Connections:
The 3000TA is a single-chassis

instrument, which has no Remote Valve Unit. Instead, the Remote Valve
connections are used as a method for directly controlling external
sample/zero/span gas valves. See Figure 3-5.