Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

23

Table 3-1: Analog Output Connections Pin Function

Pin Function
3

+ Range ID, 4-20 mA, floating

4

– Range ID, 4-20 mA, floating

5

+ % Range, 4-20 mA, floating

6

– % Range, 4-20 mA, floating

8

+ Range ID, 0-1 VDC

23

– Range ID, 0-1 VDC, negative ground

24

+ % Range, 0-1 VDC

7

– % Range, 0-1 VDC, negative ground

Alarm Relays: The nine alarm-circuit connector pins connect to

the internal alarm relay contacts. Each set of three pins provides one set
of Form C relay contacts. Each relay has both normally open and
normally closed contact connections. The contact connections are shown
in Table 3-2. They are capable of switching up to 3 amperes at 250 V ac
into a resistive load. The connectors are:

Threshold Alarm 1:

Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).

Can be configured as failsafe or non-failsafe.

Can be configured as latching or non-latching.

Can be configured out (defeated).

Threshold Alarm 2:

Can be configured as high (actuates when concentration
is above threshold), or low (actuates when concentration
is below threshold).

Can be configured as failsafe or non-failsafe.

Can be configured as latching or non-latching.

Can be configured out (defeated).

System Alarm:

Actuates when DC power supplied to circuits is
unacceptable in one or more parameters. Permanently