Operation

4.1 Introduction

Once the analyzer has been installed, it can be configured for your

application. To do this you will:

Set system parameters:

Establish a security password, if desired, requiring
Operator to log in.

Establish and start an automatic calibration cycle, if
desired.

Calibrate the instrument.

Define the three user selectable analysis ranges. Then choose
autoranging or select a fixed range of analysis, as required.

Set alarm setpoints, and modes of alarm operation (latching,
failsafe, etc).

Before you configure your 3000TA these default values are in

effect:

Ranges: LO = 100 ppm, MED = 1000 ppm, HI = 10,000 ppm.
Auto Ranging: ON

Alarm Relays: Defeated, 1000 ppm, HI, Not failsafe, Not

latching.

Zero: Auto, every 0 days at 0 hours.
Span: Auto, at 000008.00 ppm, every 0 days at 0 hours.

If you choose not to use password protection, the default password

is automatically displayed on the password screen when you start up,
and you simply press Enter for access to all functions of the analyzer.

4.2 Using the Data Entry and Function Buttons

Data Entry Buttons: The ◄►arrow buttons select options from

the menu currently being displayed on the VFD screen. The selected
option blinks.