Operation 3000TA-

EU

Teledyne

Analytical

Instruments

32

When the selected option includes a modifiable item, the ▲/▼

arrow buttons can be used to increment or decrement that modifiable
item.

The Enter button is used to accept any new entries on the VFD

screen. The Escape button is used to abort any new entries on the VFD
screen that are not yet accepted by use of the Enter button.

Figure 4-1 shows the hierarchy of functions available to the

operator via the function buttons. The six function buttons on the
analyzer are:

Analyze.

This is the normal operating mode. The analyzer

monitors the oxygen content of the sample, displays the
percent of oxygen, and warns of any alarm conditions.

System.

The system function consists of six subfunctions

that regulate the internal operations of the analyzer:

Auto-Cal setup

Password assignment

Self-Test initiation

Checking software version

Logging out.

Zero

. Used to set up a zero calibration.

Span.

Used to set up a span calibration.

Alarms.

Used to set the alarm setpoints and determine

whether each alarm will be active or defeated, HI or LO
acting, latching, and/or failsafe.

Range.

Used to set up three analysis ranges that can be

switched automatically with auto-ranging or used as
individual fixed ranges.

Any function can be selected at any time by pressing the

appropriate button (unless password restrictions apply). The order as
presented in this manual is appropriate for an initial setup.

Each of these functions is described in greater detail in the

following procedures. The VFD screen text that accompanies each
operation is reproduced, at the appropriate point in the procedure, in a
Monospaced type style. Pushbutton names are printed in Oblique type.