Trace Oxygen Analyzer

Maintenance

Teledyne

Analytical

Instruments

59

3. With one hand placed underneath the cell block ready to

catch the Micro-Fuel cell, lift up on the stainless steel gate in
front of the cell block. This releases the cell and cell holder
from the block. The cell and holder will fall out in your hand.

WARNING: Risk of electric shock high voltage exposed at

the end of enclosure!

Figure 5-1: Removing the Micro-Fuel

5.2.4 Installing a New Micro-Fuel Cell

It is important to minimize the amount of time that a Teledyne

Trace Oxygen Sensor is exposed to air during the installation process.
The quicker the sensor can be installed into the unit, the faster your TAI
O

2

sensor will recover to low O

2

measurement levels.