beautypg.com

Installation – Teledyne 3000TA-EU - General purpose trace oxygen analyzer User Manual

Page 29

background image

Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

17

Installation

Installation of the Model 3000TA Analyzer includes:

1. Unpacking
2. Mounting
3. Gas connections
4. Electrical connections
5. Installing the Micro-Fuel Cell
6. Testing the system.

3.1 Unpacking the Analyzer

Although the analyzer is shipped complete, certain parts, such as

fuses and sensors, are wrapped separately to be installed on site as part
of the installation. Carefully unpack the analyzer and inspect it for
damage. Immediately report any damage or shortages to the shipping
agent.

3.2 Mounting the Analyzer

The Model 3000TA is for indoor use in a general purpose area. It is

NOT for hazardous environments of any type.

The standard model is designed for flush panel mounting. Figure 3-

1 is an illustration of the 3000TA standard front panel and mounting
bezel. There are four mounting holes—one in each corner of the rigid
frame. The Drawings section in the rear of this manual contains outline
dimensions and mounting hole spacing diagrams.

On special order, a 19" rack-mounting panel can be provided. For

rack mounting, one or two 3000 series analyzers are flush-panel
mounted on the rack panel. See Appendix for dimensions of the
mounting panel.