Installation 3000TA-

EU

Teledyne

Analytical

Instruments

24

configured as failsafe and latching. Cannot be defeated.
Actuates if self test fails.

Reset by pressing button to remove power. Then press

again and any other button EXCEPT System to

resume.

Further detail can be found in Chapter 4, Section 4-5.

Table 3-2: Alarm Relay Contact Pins

Pin

Contact

45

Threshold Alarm 1, normally closed contact

28

Threshold Alarm 1, moving contact

46

Threshold Alarm 1, normally open contact

42

Threshold Alarm 2, normally closed contact

44

Threshold Alarm 2, moving contact

43

Threshold Alarm 2, normally open contact

36

System Alarm, normally closed contact

20

System Alarm, moving contact

37

System Alarm, normally open contact


Digital Remote Cal Inputs: Accept 0 V (off) or 24 VDC (on)

inputs for remote control of calibration. (See Remote Calibration
Protocol
below.) See Table 3-3 for pin connections.
Zero: Floating input. 5 to 24 V input across the + and – pins puts the

analyzer into the Zero mode. Either side may be grounded at
the source of the signal. 0 to 1 volt across the terminals allows
Zero

mode to terminate when done. A synchronous signal must

open and close the external zero valve appropriately. See
Remote Probe Connector. (The –C option internal valves
operate automatically.)


Span: Floating input. 5 to 24 V input across the + and – pins puts the

analyzer into the Span mode. Either side may be grounded at the
source of the signal. 0 to 1 volt across the terminals allows Span
mode to terminate when done. A synchronous signal must open
and close external span valve appropriately. See Figure 3-5