Trace Oxygen Analyzer

Installation

Teledyne

Analytical

Instruments

25

Remote Probe Connector. (The –C option internal valves
operate automatically.)

Cal Contact: This relay contact is closed while analyzer is

spanning and/or zeroing. (See Remote Calibration Protocol below.)

Table 3-3: Remote Calibration Connections

Pin Function
9

+ Remote Zero

11

– Remote Zero

10

+ Remote Span

12

– Remote Span

40 Cal

Contact

41 Cal

Contact


Remote Calibration Protocol:
To properly time the Digital

Remote Cal Inputs to the Model 3000TA Analyzer, the customer's
controller must monitor the Cal Relay Contact.

When the contact is OPEN, the analyzer is analyzing, the Remote

Cal Inputs are being polled, and a zero or span command can be sent.

When the contact is CLOSED, the analyzer is already calibrating. It

will ignore your request to calibrate, and it will not remember that
request.

Once a zero or span command is sent, and acknowledged (contact

closes), release it. If the command is continued until after the zero or
span is complete, the calibration will repeat and the Cal Relay Contact
(CRC) will close again.

For example:

1. Test the CRC. When the CRC is open, Send a zero command

until the CRC closes (The CRC will quickly close.)

2. When the CRC closes, remove the zero command.
3. When CRC opens again, send a span command until the

CRC closes. (The CRC will quickly close.)

4. When the CRC closes, remove the span command.

When CRC opens again, zero and span are

done, and the sample is being analyzed.