Thermal Conductivity Transmitter

Maintenance

Symptom Cause

Remedy

Blown fuse on input
power circuit.

Check fuse and replace if

necessary with 400 mA
fuse. See Section 4.3.

No output signal

Blown fuse





Cable disconnected
or broken

Problem with external
read-out device

A) Check fuse on signal

output circuit. Replace if
necessary with 50 mA
fuse. See Section 4.3.

B) Check fuse on output

power circuit. Replace if
necessary with 200mA
fuse. See Section 4.3.

Check power cable in
transmitter module. See
Figure 4.1.

Check external signal read-

out device connection.

Erratic output signal

Instrument out of
calibration or
improperly calibrated


Temperature
instability inside
transmitter case

Loose cable
connection




Damaged sensor

A) Recalibrate instrument

with known calibration
gases.

B) Use appropriate calibration

gas

A) Adjust heater temperature

and recalibrate. See
Section 3.6.1.

A) Check interconnection

cable.

B) Check heater cable. See

Figure 4.1.

C) Check sensor cable. See

Figure 4.1.

A) Replace sensor. Contact

factory.