Installation 2000

XTC

Figure 3-2: Mounting Dimensions for Interface Unit

Once the transmitter and galvanic isolator are securely mounted,

connect the interface cable to the transmitter. If the cable is not attached
to the safety barrier, connect the wires to the unit as described in Section
3-5. The transmitter end of the cable is fitted with a 5-pin connector.
Secure the cable to the mating connector on the transmitter.

3.4 Gas Connections

The unit is manufactured with 1/4-inch stainless steel tube fittings

for the inlet and outlet gas connections. Refer to Figure 2.2 in Section
2.3 for a recommended sample system.

1. Insert the tube into the tube fitting, and finger-tighten the nut

until the tubing in the fitting cannot be rotated freely by hand.
(This may require an additional 1/8 turn beyond finger-tight.)

2. Hold the fitting body steady with a backup wrench, and with

another wrench rotate the nut another 1-1/4 turns.

Note: Do not crimp or bend the tubing when attaching to the

transmitter.