Thermal Conductivity Transmitter

Installation

To span calibrate this instrument:

1. Connect a read-out device to the 4-20 mA output signal

terminals on the galvanic isolator module according to
Section 3-5.

2. If the transmitter cover has not been removed, unscrew the 4

cover screws and remove the top cover.

3. Allow span gas to flow through the transmitter. After a few

moments, the output device should respond to the known
concentration of gas in the span gas mixture.

4. If the read-out device indicates a value different than the

known concentration, adjust the span potentiometer (R38) as
shown in Figure 3-5 until the readout is correct. If using a
mA current read-out device, refer to Table 3-1 to determine
concentration from the output current.

5. If the zero calibration must be changed after initial span

calibration, recalibrate the span.

Table 3-2: Concentration from Output Current

Concentration

Current

Output

(mA)

0-1%

Full

Scale

0-10%

Full

Scale

0-50%

Full

Scale

90-100%

Full

Scale

0-100%

Full

Scale

4 0 0

0 90 0

5.6 0.1 1 5 91 10

7.2 0.2 2 10 92 20

8.8 0.3 3 15 93 30

10.4 0.4 4 20 94 40

12.0 0.5 5 25 95 50

13.6 0.6 6 30 96 60

15.2 0.7 7 35 97 70

16.8 0.8 8 40 98 80

18.4 0.9 9 45 99 90

20 1.0 10 50 100 100