Range limits and indication – Meriam M2110F SMART FLOW GAUGE User Manual
Page 35
35
Range Limits and Indication
Sensor Hard
Soft
Certified
Soft
Hard
Type
Under Under
Range
Over
Over
Range Range
(F.S.)
Range
Range
DN
Under
-20%
1
-20 to 0%
0 to 100%
100 to
120%
Over
120%
D
I
Under
-20%
2
-20 to 0%
0 to 100%
100 to
120%
Over
120%
G
I
Under
-20%
3
-20
3
to
0%
0 to 100%
100 to
120%
Over
120%
C
I
NA
4
NA
4
0 to 100%
100 to
120%
Over
120%
A
I
NA
5
NA
5
0 to 100%
100 to
120%
Over
120%
Notes:
1
DN units will measure to -20% (or -10 PSID, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
2
D
I
units will measure to -20% (or -150 PSI, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
3
G
I
units will measure to -20% (or -10 PSIG, whichever is greater
pressure relative to absolute zero) and +120% of full scale pressure.
Calibration certification is for 0 – F.S. (Full Scale Only).
4
C
I
units will measure to negative barometric pressure (typically -14.7
PSIG) and to +120% of full scale pressure. Calibration certification is for
-14.5 PSIG to F.S. (Full Scale Only).
5
A
I
units will measure to +120% of full scale pressure. Calibration
certification is for 0 PSIA to F.S. (Full Scale Only).
Hard Over & Under:
Display will read "OP" (Under -20% or Over 120%).
Soft Over & Under:
Display will still read out at (± 20%).