12

5.1 PROGRAMMABLE REGISTER LIST

All M2110F Flow Gauges have programmable registers that allow the gauge to be configured to fit the flow
measurement application. The programmable registers are numbered P0 through P10 and each register controls a
specific aspect of the gauge’s performance.

During any register editing operation the display will return to the measure mode display after one minute of keypad
activity. This will ensure that the gauge will be available for flow measurement observation even if it was
unintentionally left in the program display mode.

P#

Name

Description

Value Range

Notes

P0

Lockout Code

Lockout for security.

00 to 99

00 = Disabled.
Pg. 14

P1

Time-out

Automatic shutoff in minutes of
keypad inactivity.

0 (disabled), 1, 2, 5,
10, 15, 25

Battery models only.
Pg. 14

P2

Damp Rate

Exponential damping time in seconds.

0.1, 0.2, 0.5, 1, 2, 5,
10, 15, 25, 50

0.1 = No Damping.
Pg. 15

0) inH2O display

Std. InH2O @ 20

ºC

Pg. 15

1) Laminar Flow
Element

LFE InH2O @ 4

ºC

P3 Measurement

Mode

Identifies the primary element
characteristic (Square Root or Linear)

2) Square Root
Element

SQRT InH2O @ 60

ºF

P4 Low

Flow

CutOff

Minimum display value that will be
viewed on the LCD and Serial I/O.

-19999 to Max. Full
Scale

Decimal flashes when
activated.
Pg. 16

SetPoint Model: defines which relay
outputs are active.

0 = Disabled.
1 = SET1 only.
2 = SET2 only.
3 = Both enabled.

P5 SetPoint

Options

Current Loop Model: defines the
status of the 4 - 20 mA outputs.

0 = 4-20 disabled
1 = 4-20 enabled

Setpoint and Current
Loop models only.
Pg. 17

P6

SET1

Controls SET1 relay or 4.00 mA
value.

–20% to +120% FS

User defined value.
Pg. 17

P7

SET2

Controls SET2 relay or 20.0 mA
value.

–20% to +120% FS

User defined value.
Pg. 17

P8

Deadband

Sets the amount of deadband in
percent of full scale for relays.

0 (disabled), 0.1, 0.2,
0.5, 1, 2, 5, 10%

SetPoint model only.
Pg. 19

P9 B-Coefficient

/

Flow Constant
Scaling




Engineering
Units

The scaling constant required to
calibrate the flow measurement
equation.




After the entry of numeric data, access
to the units selection is allowed.

Mantissa range
+/- 9,999
Exponent Range
E+9 to E-9
Total Range
0 to 2250

A valid P9 setting is
required for P3=1 or 2.




User can scroll
through possible
engineering units.
Pg. 20

P10 C-Coefficient

Scaling

The scaling constant required to
calibrate the flow measurement
equation.

Mantissa range
+/- 9,999
Exponent Range
E+9 to E-9
Total Range
+/- 2.0

P10 required for LFE
devices.
P3 = 2 only.
Pg. 21