beautypg.com

Meriam M2110F SMART FLOW GAUGE User Manual

Page 31

background image

31

14.1 ELECTRICAL CONNECTIONS

Setpoint models come with a terminal block interface while the Current Loop model comes with an eight-pin
circular connector/cable interface. Note: The cable is supplied with the Current Loop model.

14.1.1 Power Supply Options

The M2110F SMART GAUGE can be supplied in three possible hardware configurations.


1. Battery Model ..………….Unit powered by internal batteries

NO EXTERNAL POWER SUPPLY REQUIRED


2. Setpoint Model …………..Unit includes two programmable relay outputs

EXTERNAL POWER SOURCE: 24 VDC, 115 VAC or 230 VAC


3. Current Loop Model …….Unit includes 4-20 mA output

EXTERNAL POWER SOURCE: 24 VDC

14.1.2 Setpoint Model Interface Connections

Terminal

Description

Function

Notes

1

SET1 / NO

Relay1 output

Normally open

2

SET1 / C

Relay1 output

Common

3

SET1 / NC

Relay1 output

Normally closed

4

SET2 / NO

Relay2 output

Normally open

5

SET2 / C

Relay2 output

Common

6

SET2 / NC

Relay2 output

Normally closed

7

No Connection

None

8

RS232 Common

RS232 interface common connection

0 VDC

9

RS232 Tx

RS232 transmit output connection

For Serial Menu

10

RS232 Rx

RS232 receive input connection

For Serial Menu

A

Frame Ground

Earth Connection

B

GND

AC Ground

DC POWER SUPPLY CONNECTIONS

C

0 VDC

Common

Connect only if not using AC inputs

D

+24 VDC IN

+24 VDC Input Nominal

Connect only if not using AC inputs

AC POWER SUPPLY CONNECTIONS
Reconnectable 115/230 Primary

115 VAC
INPUT

230 VAC
INPUT

E

Reconnectable Transformer Primary

Jumper E to H

NC

F

Reconnectable Transformer Primary

Jumper F to J

NC

H

L2

Reconnectable Transformer Primary

(See E above)

L2 / 230 VAC

J

Reconnectable Transformer Primary

(See F above)

Jumper J to K

K

N

Transformer Primary Fused Input

Line Neutral

L

L1 / L

Transformer Primary Fused Input

L 115 VAC

L1 / 230VAC


Note: See Appendix for wiring specifics.