5 programming – Meriam M2110F SMART FLOW GAUGE User Manual
Page 11
11
5 PROGRAMMING
IMPORTANT:
Programmable registers above P0 cannot be adjusted unless the correct lockout code is entered or the lockout code is
disabled.
PROGRAM MODE ACCESS
no
[1] [2] [3] [5] [6] [8]
yes
no
no
yes
yes
Notes
[1]
Enter program register selection mode
[2]
Select register number to change
[3]
View the value of the register currently selected
[4]
Do you want to change the value? If not go to [6]
[5]
Increase or decrease the value of the blinking digit, enter PRGM after each digit adjustment is made
[6]
After modifying the final digit, PRGM will “store” the new value into memory
[7]
Do other programmable registers need to be changed? If so, return to [2]
[8]
Exit the program select mode and return to measurement display mode
Key Functions
Backspace : This key can be used to abort data entry and escape back to the measure mode display
Program : Enters data for storage or mode change
Up Arrow : Increments menu selections or digits being edited
Down Arrow : Decrements menu selections or digits being edited
[4]
Change
?
[7]
Change
?
Press
Prgm
Press
Press
Prgm
Press
Prgm
Press
ON/
OFF
Press