4.2.4
PRGM or ENTER
Mode of
Operation
PRGM
Document Symbol: (PRGM/Enter
→)
Measure Mode
Selects Program Mode, which allows access to the programmable registers P0-P10.
Program Mode
Program Mode (P# displayed): Opens the selected register for editing.
Data Edit Mode: When the desired value is selected, pressing the PRGM/ENTER
→
key accepts and stores the value.
When editing a multi-digit value, each key press will accept the current digit and proceed
to the next until finally accepting the complete value and closing the register.
4.2.5
ZERO FUNCTION
Mode of
Operation
ENG
UNITS
+
Document Symbol: (Up
↑ and Down↓)
Measure Mode
In Measure Mode, pressing the Up
↑ and Down↓ keys at the same time resets the zero
reference of the gauge (see page 16).
Program Mode
In Program Mode, for convenience, this function will reset to P0.
Program Mode (P# displayed): Function will reset the register index to P0.
Data Edit Mode: After opening a register for editing, this function will reset the
register’s data value to default.
Note that even when using this “reset” function, the PRGM/Enter
→ key must be
pressed to accept the new value. This allows for further editing after reset of the value.